7100 Series Hipot / Insulation Tester

MODEL 7151 7152 7160 7162
AC WITHSTAND VOLTAGE
Output Rating 5KVAC/12mA
  Range Resolution Accuracy
Output Voltage, KVAC 0-5.00 0.01 ±(2% of setting + 5V)
Output Frequency 50Hz/60Hz ±100ppm, User Selection
Output Waveform Sine Wave ,THD.<2% (Resistive Load), Crest Factor = 1.3 - 1.5
Output Regulation ± (1% of output + 5V), From no load to full load
SETTINGS
Hi-Limit AC Current, mA 0.10 - 12.00 0.01 ±(2% of setting + 2counts)
Lo-Limit AC Current, mA 0.00 - 12.00 0.01 ±(2% of setting + 2counts)
Ramp Time, second 0.1 - 999.9 0.1 ±(0.1% + 0.05sec)
Dwell Time, second 0, 0.2 - 999.9
(0=continuous)
0.1
Arc Detection 0, 1 - 9 ranges (0=OFF, 9 is the most sensitivity)
DC WITHSTAND VOLTAGE (7160,7162 ONLY)
Output Rating 6KVDC/5mA
Output Voltage, KVDC 0 - 6.00 0.01 ±(2% of setting + 5V)
Output Ripple <5%(6KV/5mA at Resistive Load)
SETTINGS
Hi-Limit DC Current, mA 0.02 - 5.00 0.01 ±(2% of setting + 2counts)
Lo-Limit DC Current, mA 0.00 - 5.00 0.01 ±(2% of setting + 2counts)
Ramp Time, second 0.1 - 999.9 0.1 ±(0.1% + 0.05sec)
Dwell Time, second 0, 0.2 - 999.9
(0=continuous)
0.1
Arc Detection 0, 1 - 9 ranges (0=OFF, 9 is the most sensitivity)
Discharge Time < 200 msec
INSULATION RESISTANCE (7152,7162 ONLY)
Output Rating 1KVDC/1000MΩ
Output Voltage, KVDC 0.10 - 1.00 0.01 ±(2% of setting + 3V)
SETTINGS
Hi-Limit Resistance, MΩ 0, 1 - 1000 (0=OFF) 1 ±(3% of setting+2counts)at > 500VDC
Lo-Limit Resistance, MΩ 1 - 1000 1 ±(7% of setting +2counts)at <500VDC
Delay Time, second 0, 0.5 - 999.9
(0=continuous)
0.1 ±(0.1% + 0.05sec)
MEASUREMENT
AC/DC Voltage, KV 0 - 6.00 0.01 ± (1.5% of reading ) ≥ 500V
± (1.5% of reading + 1 count) < 500V
DC Voltage, KV (IR only) 0.10 - 1.00 0.01 ±(1.5% of reading + 3V)
AC Current, mA 0.10 - 12.00 0.01 ±(2% of reading + 2counts)
DC Current, mA 0.02 - 5.00 0.01 ±(2% of reading + 2counts)
Resistance, MΩ 1 - 1000
(Auto Range)
0.001
0.01
0.1
±(3% of reading+2counts)at > 500VDC
±(7% of reading +2counts)at <500VDC *1
GENERAL
Input Voltage AC 115/230Vac±15%, 50/60Hz ± 5%, Fuse 3.15A Fast
PLC Remote Control Input : Test , Reset
Output : Pass , Fail , Processing
Memory 10 Sets Store/Recall of each Memory, 4 Test Mode Selectable (W, I, W~I, I~W Only 7152, 7162)
Display 16×2 LCD with back light
Key Lock To prevent unauthorized alteration of the test Parameters
Calibration Build-in software and external calibrated meters
Environment 0-40℃, 20-80%RH
Dimension/Net Weight 280mm(W)×89mm(H)×300mm(D)/ 8.4Kg
STANDARD ACCESSORIES
Power Cord(10A) ×1
Fuses ×2 (Including a spare contained in the fuse holder)
Hipot Test Lead ,1.5m(1101) ×1
Hipot Return Lead ,1.8m(1102) ×1
*1 If the value of V/R is lower than 3.5uA for IR function, 7152 and 7162 need to be warmed for one hour
*product specifications are subject to change without notice.